29 30
5. Result range: 20 mg/dL ~600 mg/dL(1.1 mmol/L ~33.3mmol/L)
6. Glucose Units: Either mg/dL or mmol/L (USA version preset
mg/dL only)
7. Memory : 500 blood glucose tests
8. Display: LCD display with backlight
9. Power source: DC 3V (AAA×2 alkaline batteries)
10. Battery Life: Approx. 500 normal tests
1
1. Storage Condition: Blood glucose Monitoring system (meter
and test strips): 39 86 , Humidity80%RH.
12. Storage Condition (meter): -40 131 , Humidity80%RH
13. Operating conditions: 50 104 , Humidity:25%~80%RH
14. Blood Source: Fresh capillary whole blood
15. Blood Volume: Minimum 0.7 micro liter.
16. Use life: five years
NOTE: Factors affecting the life of the meter
The factor
The aging
of the strip
connector
The clou-
ding of the
lens
The aging
of the el-
ectronic
components
The indication
The meter cannot be
turned on when In-
serting the test strip
several times.
The content displayed
on the lens cannot be
recognized clearly.
The results are not
accurate when the
test strips are in
normal mode.
Do this:
please contact local cus-
tomer services or place
of purchase for assistance,
or toll free service hotline:
1-877-342-2385
please contact local cus-
tomer services or place
of purchase for assistance,
or toll free service hotline:
1-877-342-2385
please contact local cus-
tomer services or place of
purchase for assistance,
or toll free service hotline:
1-877-342-2385
The blood Monitoring System is not intended for use on neo-
nates.
The blood Monitoring System is not intended for use on
artery blood, neonates’ serum and plasma.
The Blood Glucose Meter should be used with “CLEVER
CHOICE” test strip.
The system can be used up to an altitude of 10744 feet.
The following substances at levels greater than normal or
therapeutic levels may cause significant interference (affect
the result by greater than 10%) resulting in an inaccurate
result: ascorbic acid, uric acid, hemoglobinacetaminophen,
Dopamine, L-dopa, L-dopa and Tolbutamide etc. These sub-
stances do not affect test results in normal concentration but
may affect test results in high concentration. Do not use hae-
molysis sample, icterus sample or high lipemia samples.
The following compounds, when determined to in excess of
their limitation, may produce inaccurate results:
Compound
Ascorbic acid
Uric acid
Acetaminophen
Bilirubin
Dopamine
L-dopa
Methyldopa
Tolbutamide
Triglycerides
Hemoglobin
Limitation
2mg/dL
10mg/dL
5mg/dL
15mg/dL
0.03mg/dL
0.45mg/dL
0.75mg/dL
24mg/dL
2000mg/dL
250mg/dL