31 32
Patients undergoing oxygen therapy may yield falsely lower
results.
Not for use for patients in a hyperglycemic-hyperosmolar
state, with or without ketosis.
Not for use on critically ill patients.
Not to be used for patients who are dehydrated, hypertensive,
hypotensive or in shock.
Very low (less than 20%) or very high (more than 60%) red
blood cell count (hematocrit) can lead to incorrect test results.
If you do not know your hemarocrit level, please consult your
health care provider.
MESSAGE
WHAT IT MEANS
Blood glucose level is
lower than 20mg/dl
(1.1mmol/L)
Blood glucose level is
higher than 600 mg/dl
(33.3mmol/L)
Meter is ready to test
control solution under
the CTL state.
Battery power is low.
ACTION
The message indicates
very low blood sugar. You
should consult with your
healthcare professional.
This indicates sever
hyperglycemia (high
blood glucose).
You should seek imme-
diate medical assistance.
Please place a drop of
control solution onto the
test well of test strip.
Please change the
battery.
MESSAGE
WH AT IT MEANS
Use a used test strip.
Error massage indicates
that you may remove the
strip after applying blood
to the absorbent hole.
Problem with the meter.
Environmental tem-
perature is lower than
10 (50 ) or higher
than 40 (104 )
ACTION
Please use a new strip
and re-test.
Try again with a new
test strip.
Re-test with a new test
strip. If the problem is
still unsolved, please
call the Customer Care
Line for help.
Please test at an envi-
ronmental temperature
Display remains
blank after the
test strip has
been inserted
into the Meter.
1. Battery may be
installed improperly.
2. Battery power is
too low for use.
3. Too much time has
passed between
inserting the Test
Strip and perfor-
ming the test.
1.Install the Battery
correctly.
2.Replace with new
Battery.
3.Reinsert the Test
Strip into the
Meter.
Vista de página 16